TECHNICAL GUIDE

PlateLoc seal selection guide

Check web handling, seal integrity, removal, storage and solvent resistance when selecting a roll film.

Check the roll and equipment first

Confirm roll width, core, winding direction and loading method. The catalog lists LX11031R as a 115 mm × 400 m roll and recommends 150–180°C for 3–5 seconds. During trial runs, check web centering, tension, cutting and take-up.

Four performance checks

CheckMethodRecord
Seal integrityInspect every well rim for continuous contact and recheck under real storage conditions.Lifted edges, channels, damage and leakage
Peeling and piercingUse the intended access method and check for tearing or residue.Removal consistency, piercing and residue
Evaporation and storageCompare mass or volume before and after storage using the same plate and fill volume.Time, temperature and edge-well differences
Chemical compatibilityTest the actual DMSO or other solvent system against the film, adhesive and seal.Solvent, concentration and contact time

Five-step validation checklist

  1. Record the current plate, instrument, film, sample and storage conditions.
  2. Seal a test batch using the candidate film's recommended settings.
  3. Inspect every well, then test peeling or piercing.
  4. Measure evaporation and leakage under real storage and transport conditions.
  5. Document final settings and lots before a small continuous run.

The catalog lists LX11031R as a cross-reference to Agilent 24214-001. Confirm selection against the actual plate and instrument results.